Fisher & Paykel MR850 Humidifier with Heated Wire Circuit and Temperature Probe

Fisher & Paykel MR850 Humidifier with Heated Wire Circuit and Temperature Probe

Inquire Now

Clinical-Grade Humidification for Every Stage of Respiratory Care

The Fisher & Paykel MR850 is a heated respiratory humidifier engineered for use in hospital intensive care units. It warms and humidifies breathing gases delivered to patients on mechanical ventilation, positive pressure support, or supplemental oxygen therapy — preserving the natural mucociliary function of the airway and reducing the risk of secretion retention and infection.

What sets the MR850 apart is its dual-mode intelligence. A single button toggles between Invasive Mode (37 °C, 44 mg/L — body temperature with full saturation) for intubated patients, and Non-Invasive Mode (31 °C, 32 mg/L) for patients on mask-based therapy. Built-in ambient sensing algorithms continuously adjust heater plate output to compensate for changes in room temperature, maintaining consistent humidity delivery without manual intervention.

The system ships as a complete, ready-to-use package comprising the MR850 heater base, a heated-wire adapter, and a reusable temperature and flow probe (Ref. 900MR86X). The temperature probe connects to the blue socket on the heater base and feeds real-time gas temperature data from both the chamber outlet and the patient airway — enabling precise closed-loop power control across both the heater plate and the heated circuit wires.

Suitable for adult, pediatric, and neonatal patients, the MR850 supports the full Fisher & Paykel Respiratory Care Continuum — from invasive ventilation and NIV through to Optiflow high-flow oxygen therapy. Its service life of 7 years and straightforward surface-cleanable design make it a durable, cost-effective choice for Indian hospitals and critical care facilities.

Package Includes

MR850 heater base + heated-wire adapter (Ref. 900MR8XX) + temperature & flow probe (Ref. 900MR86X). Humidification chamber and breathing circuit ordered separately.

Authorized Source in India

Procured through Shah Trading Corporation — ensuring genuine F&P Healthcare products, local support, and service documentation compliance.

Key Features

Dual-Mode Operation - One-button switch between Invasive (37 °C / 44 mg/L) and Non-Invasive (31 °C / 32 mg/L) modes — covering the full range of ventilatory support with clinically targeted humidity levels.

Closed-Loop Temperature Control - The temperature probe provides continuous feedback from the chamber and airway, allowing the MR850 to automatically adjust heater plate power for stable, precise gas delivery.

Heated Wire Circuit Integration - The heater-wire adapter powers both the inspiratory and expiratory limbs of the breathing circuit, keeping water vapour above dew point to minimize condensate ("rain-out") in the tubing.

Clinically Based Alarm System - Medium-priority alarms trigger immediately at ≥ 41 °C displayed temperature or ≥ 43 °C airway probe reading. Automatic heater shutdown with flashing display protects patients from overheating.

Ambient Environment Compensation - Built-in algorithms detect changes in room temperature and automatically adjust output — preventing inconsistent humidity delivery without clinician intervention.

Universal Patient Suitability - Supports adult, pediatric, and neonatal patients. Compatible with the full F&P Respiratory Care Continuum — from ICU invasive ventilation to Optiflow high-flow nasal therapy.

Flow-Sensing Intelligence - Integrated flow sensing up to 60 LPM (invasive) and 120 LPM (non-invasive) allows the system to optimize humidity output dynamically, reducing nuisance alarms during variable ventilation.

Auto Standby & Power Safety - Auto-standby mode reduces unnecessary power use between patient connections. Thermal cutout at 118 ± 6 °C on the heater plate ensures a fail-safe hardware protection layer independent of software.

Technical Specifications

Parameter Value / Range
Operating Modes Invasive (37 °C, 44 mg/L) | Non-Invasive (31 °C, 32 mg/L)
Invasive Mode Temp. Range 35.5 °C – 40 °C (adjustable)
Non-Invasive Mode Temp. Range 31.0 °C – 36.0 °C (adjustable)
Display Temperature Range 10 °C – 70 °C
Temperature Probe Accuracy ± 0.3 °C (in 25 – 45 °C range)
Gas Temperature Measurement Accuracy ± 2 °C
High Temperature Alarm — Displayed ≥ 41 °C (Medium Priority, immediate)
High Temperature Alarm — Airway Probe ≥ 43 °C (Medium Priority, immediate)
Flow Range — Invasive Mode Up to 60 LPM
Flow Range — Non-Invasive Mode Up to 120 LPM
Heater Plate Capacity 150 W at nominal mains voltage
Heater Plate Thermal Cutout 118 ± 6 °C
Heater Wire Supply 22 V~, 60 W max
Total Max. Power Consumption 210 W
Supply Voltage (India — AXX variant) 230 V~, 50/60 Hz, 1.0 A max
Dimensions (without chamber) 140 × 173 × 135 mm
Weight (without chamber) 2.8 kg
Weight (with filled chamber) 3.1 kg
Warm-Up Time < 30 minutes
Recommended Ambient Temperature 18 °C – 26 °C
Temperature Probe Ref. No. 900MR86X (reusable; not for autoclave)
Heater-Wire Adapter Ref. No. 900MR8XX
Cleaning Agent (Heater Base) Isopropyl alcohol or standard dishwashing detergent (damp cloth only — do not immerse)
Service Life 7 years (designed for daily hospital use)
Intended Setting Hospital ICU / Critical Care
Patient Population Adult, Pediatric, Neonatal
Mounting Options Flat surface, ventilator rail, pole stand (via Mounting Bracket 900MR303)
View full details
Frequently Asked Questions

We run both ventilated and NIV patients in our ICU — does the MR850 need to be switched between units, or can one device handle both?

A single MR850 unit handles both scenarios without any hardware change. Pressing the mode button toggles between Invasive Mode (37 °C / 44 mg/L for intubated patients) and Non-Invasive Mode (31 °C / 32 mg/L for mask-based therapy). This means one device can follow a patient across therapy transitions — from NIV to intubation — reducing equipment cost and setup time.

What exactly does the temperature probe do, and is it reusable or single-use?

The temperature probe (Ref. 900MR86X) contains two sensing points — a Chamber Probe and an Airway Probe. They feed real-time temperature data back to the heater base, which uses this information to fine-tune power output to both the heater plate and the heated circuit wires. This closed-loop feedback is what makes the MR850's delivery accurate to ± 0.3 °C. The probe is reusable and can be cleaned with isopropyl alcohol — it must not be autoclaved or immersed in liquid.

Is the MR850 safe for use with neonates? Our NICU runs very sensitive patients.

Yes — the MR850 is explicitly designed for adult, pediatric, and neonatal patients. The temperature alarm triggers immediately at 41 °C (display) or 43 °C (airway probe), with automatic shutdown of both the heater plate and heater wire. For neonates, this rapid, hardware-level overheat protection is critical. The system also supports low flow rates appropriate for neonatal circuits. Fisher & Paykel Healthcare has published clinical evidence supporting its use in neonatal populations across prematurity and respiratory distress conditions.